AAS Peak


The AAS is fully PC controlled with modern and compact design. They have flexible configuration and different accessories to meet variable requirements. Advance optical system can guarantee stable and reliable performance. Unique and intelligent gas control module can ensure the device safety. The AAS can test more than 70 elements including metallic and non-metallic elements, which can be used widely in petrochemistry, metallurgy, mining, geology, material analysis, food and pharmacy.

Flame & Graphite Furnace AAS Specification
Model AA6018 AA6028
Optical System Optical Type Single beam Double beam
Monochromator Czerny-Turner type with focal lengths at 350mm
Wavelength Range 190-900nm
Grating Holographic diffraction grating with 1800 lines/mm
Slits Automated slit selection 0.1, 0.2, 0.4, 0.7, 1.4, 2.0 nm
Wavelength Accuracy ±0.2 nm ±0.15 nm
Wavelength Repeatability ±0.1nm ±0.05nm
Lamp Holder Automated 8-lamp turret
Detector Wide range UV sensitive photomultiplier tube
Photometric
Performance
Reading Mode Transmittance(T), Absorbance(A), Concentration(C)
Photometric Range 0-125%, -0.1-3.00A
Static Baseline Drift (Cu) ?0.003A/30 min ?0.002A/30 min
Dynamic Baseline Drift (Cu) ?0.005A/15 min ?0.004A/15 min
Background Correction Deuterium (D2) Background Correction and
Self-absorption Background Correction
Flame Analysis Characteristic Concentration Cu?0.035ug/ml Cu?0.02ug/ml
Detection Limit Cu?0.006ug/ml Cu?0.004ug/ml
Repeatability (Cu) RSD?0.6% RSD?0.5%
Acetylene Flow Control Auto 12-level Auto 12-level
Air Flow Control Auto 4-level Auto 4-level
Nebulizer High efficiency glass
Burner Adjustment Automatic up and down
Burner Head Titanium alloy 100mm burner for air/acetylene operation
Premix Chamber Corrosion resistant
Safety Protection Flame sensor, gas leak sensor, gas pressure sensor
Graphite
Furnace
Analysis
Characteristic Concentration Cd?0.6pg , Cu?20pg Cd?0.5pg , Cu?20pg
Detection Limit Cd?1.0pg , Cu?25pg Cd?0.8pg , Cu?25pg
Repeatability (Cu) RSD?3% RSD?2.5%
Temperature Range Ambient -3000? Ambient -3000?
Ramp and holding time 1-255s 1-255s
Heating Rate Maximum linear heating rate 2000?/s
Atomization Heating Method Light-controlled heating, Time-controlled heating, General heating
Gas Flow Rate Inside Tube 1 L/min 1 L/min
Gas Flow Rate Outside Tube 4-speed adjustable (0, 50, 200, 250ml/min)
Safety Protection Water flow sensor, Argon pressure sensor, Furnace body temperature sensor, Graphite tube installation self-inspection.
Data Processing Software RGWIN AAS software package
Analytical Method Flame AA, Flame AE, Hydride Generation
Concentration Calculation
Method
Standard curve(6 linear/non-linear fitting), Standard Addition, Interpolation
Repeated Measurement Times 1-30 times, calculate the average, show SD and RSD value of absorbance and concentration
QA/QC Function Correlation coefficient, SD & RSD of Abs and calculate SD & RSD of concentration
Report Printing Parameter printing, resulting printing
Communication Port Standard RS-232
Others Computer External(excluded)
Dimension & Weight 850*650*550MM(W*D*H), 100Kg
Power Supply 110/220V, 50/60Hz
Working Environment Temperature 10 ~ 35 °C, Humidity?85%